  • ANALYTICAL MODES:
    • Manual, Predilute, Capillary
  • TOTAL PARAMETERS:
    • 33: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW-CV, PDW-SD, PCT, P-LCR, P-LCC, Neu%, Lym%, Mon%, Eos%, Bas%, Neu#, Lym#, Mon#, Eos#, Bas#, ALY%*, LIC%*, ALY#*, LIC#*, cWBC#*, NRBC%*, NRBC#*
  • PRINCIPLE OF MEASUREMENT:
  • RBC/PLT/WBC: Electrical Impedance
  • DIFF: 3 Angle Laser Flow Cytometry
  • HGB: Cyanide Free Colorimetry
  • MCV: Measured
  • HCT: Calculated
  • GRAPHICS:
    • 3 Histograms (WBC/RBC/PLT)
    • 4 Scatterplots (DIFF x3, BASO)
  • SAMPLE VOLUME:
    • Whole Blood : 15 μL
    • CBC Only : 11 μL
    • Pre-diluted : 20 μL
    • Capillary : 20 μL

  • LINEARITY RANGE:
    • WBC (x 10^3/L): 0 – 300
    • RBC (x 10^6/L): 0.00 – 8.50
    • Hb (g/dL) : 0 – 25.0
    • HCT (%): 0 – 67
    • PLT (x 10^3/L) : 0 – 3000
  • CALIBRATOR:
    • ELite H Cal (3mL)
    • Open Vial Stability at 2-8°C: 7 Days
  • TRI-LEVEL CONTROLS:
    • Erba H5 CON L, N, H (3mL)
    • Open Vial Stability at 2-8°C: 14 Days
  • THROUGHPUT:
    • 60 Tests/Hr
  • QC (L-J, X-BAR):
    • Yes
  • DATA STORAGE:
    • 50,000 Results with Graphs

  • DIMENSION (mm):
    • 364 x 498 x 431
  • WEIGHT (Kg):
    • 26.5
  • REAGENTS:
    • Erba Dil (20L) ErbaLyse 1 (200mL) Erba Lyse 2 (500mL) Elite H Clean (50mL)
  • INTERFACES:
    • 4 USB + 1 LAN Port
  • OPERATING ENVIRONMENT:
    • 15-30°C
    • Atmospheric pressure
    • 70kPa~106kPa
  • POWER REQUIREMENT:
    • A.C.100-240V; 50/60 Hz;
    • ≤200VA